Model 924 Toxic Air Sampler
The Xonteck Model 924 Toxic Air
Sampler is designed for unattended field use to collect ambient air
samples for laboratory analysis of toxic compounds.
The sampler is modular in design and consists of a control unit, pump
box assembly, rain shield, and sampling head mount. The sampler can
accept from one to four sampling channels that include 37/47mm filter
for Cr6 and other particulates, Carbonyl with optional pre-filter,
Sorbent tube with upstream actuation valve, and Canister (VOC).*
Features
● Canister, Carbonyl, 37/47mm Filter (Cr6), and
Sorbent tube multi- configurable
● Weather proof four-channel stand-alone sampling station
● Corrosion resistant stainless steel and brass fittings
● Integrated computer with printer and touch-screen display
● Temperature controlled heater and fans for cold or hot weather
operation
● Modular design for ease of assembly, installation, operation and
service
|
 |
Specifications |
Sample Flow
Control: Porter Model 201, 202 Mass Flow Controller, typically
0-2SLPM, 0-30SLPM or customer specified, accuracy and linearity
to ±1% of full scale.
Display: touch screen, TFT color graphics, white LED backlight,
5.7” diagonal, 320 x 240 pixel density.
External Sample Pump: Gast 1023 Series Motor Mounted Rotary
Vane, max vacuum 26” Hg, open flow 10.0 CFM.
Digital Vacuum Transducers: -30”Hg to 0psig, ±0.5 % of Full
Scale (BFSL).
|
Integrated
Temperature Control: Designed into PC circuit for enclosure heat
control via 100 Watt silicone rubber heater.
Temperature Display: enclosure temperature displayed, Honeywell
RTD sensor, -40 °C to 150 °C, 1 °C resolution.
Power: 115V/60Hz; 220V/50Hz(optional).
Dimensions: 7’(H)×1’10”(W) ×1’(D)
Weights:
Sampler (pump not included): 128lbs
Sample Pump: 52lbs |
|
*The sampler can accept any combination of sampling
channels with the exception of the canister channel. The sampler can
only accept one canister channel. |