Model 801
Chemiluminescence NO-NO2-NOx Analyzer

Principle of Operation

Chemiluminescence corresponds to an oxidation of NO molecules by O3 molecules: NO + O3= NO2* + O2. The return to a fundamental electronic state of the excited NO2* molecules are made by luminous radiation in a 600-1200 nanometers spectrum: NO2*NO2 + hv, Where:h- planck constant, v- frequency of photons emitted. The radiation measurement is made by a photo-multiplier. The electrical signal it delivers is amplified and digitized for treatment by the microprocessor. Nitrogen dioxide (NO2) must first be transformed into NO before it can be measured using the luminescent reaction. NO2 is converted to NO by a molybdenum converter heated to about 200 ℃

Features

● LCD touch screen

● Modular design

● Network oriented

● Auto light source control

● Auto alarm functions.

● Flexible calibration options

● Advanced digital processor

Specifications

Method: Chemiluminescence

Range: 0-500ppb

Reading error: 0.5%FS

Zero noise: ≤0.25ppb

Lower detectable limit: 0.5ppb

Repeatability: <1%

Linearity: <±1%FS

Response time: <120s

Weights: 19kgs

Zero drift:<0.5ppb/24h

Span drift:<±1.5% FS/24h

Conversion efficiency: >98%

Flow Rate: 500±10%mL/min

Digital output: RS232/RS485,USB/Ethernet

Analog output: 0-5V, 4-20mA

Resolution: 0.1ppb,0.01ppb optional

Operation temperature: 0-40℃

Dimension: 430*570*225mm